Search results for "Aspiration catheter"

showing 2 items of 2 documents

Necessary Catheter Diameters for Mechanical Thrombectomy with ADAPT

2017

BACKGROUND AND PURPOSE: Large-bore catheters allow mechanical thrombectomy in ischemic stroke by engaging and retrieving clots without additional devices (direct aspiration first-pass technique [ADAPT]). The purpose of this study was to establish a model for minimal catheter diameters needed for ADAPT. MATERIALS AND METHODS: We established a theoretic model for the calculation of minimal catheter diameters needed for ADAPT. We then verified its validity in 28 ADAPT maneuvers in a porcine in vivo model. To account for different mechanical thrombectomy techniques, we factored in ADAPT with/without a hypothetic 0.021-inch microcatheter or 0.014-inch microwire inside the lumen of the aspiration…

CathetersMechanical ThrombolysisSwine030218 nuclear medicine & medical imaging03 medical and health sciences0302 clinical medicinemedicine.arterymedicineInner diameterAnimalsRadiology Nuclear Medicine and imagingAspiration catheterInterventionalbusiness.industryPredictive valueMechanical thrombectomyStrokeCatheterTreatment OutcomeMiddle cerebral arteryNeurology (clinical)Internal carotid arterybusinessTheoretic model030217 neurology & neurosurgeryBiomedical engineering
researchProduct

Under Pressure: Comparison of Aspiration Techniques for Endovascular Mechanical Thrombectomy.

2018

BACKGROUND AND PURPOSE: Blood flow should be interrupted during mechanical thrombectomy to prevent embolization of clot fragments. The purpose of our study was to provide a handy overview of the most common aspiration devices and to quantify their flow characteristics. MATERIALS AND METHODS: We assessed volumetric flow rates generated by a 60-mL VacLok vacuum pressure syringe, a Pump MAX aspiration pump, and a Dominant Flex suction pump connected to the following: 1) an 8F long sheath, 2) an 8F balloon-guide catheter, 3) an ACE 64 distal aspiration catheter, and 4) an AXS Catalyst 6 Distal Access Catheter. We used a water/glycerol solution, which was kept at a constant temperature of 20°C (…

Cathetersmedicine.medical_treatmentVacuum pressureSuctionBalloonModels Biological030218 nuclear medicine & medical imaging03 medical and health sciences0302 clinical medicinemedicineHumansRadiology Nuclear Medicine and imagingEmbolizationSyringeThrombectomyAspiration catheterInterventionalbusiness.industrySyringesEndovascular ProceduresBlood flowMechanical thrombectomyCatheterNeurology (clinical)business030217 neurology & neurosurgeryBiomedical engineeringAJNR. American journal of neuroradiology
researchProduct